Maintenance
122 Instructions for use Babylog 8000 plus SW 5.n
Inspection
Perform inspections at regular intervals and
observe the following specifications.
Safety checks
The safety checks are no substitute for the
preventive maintenance measures (including
preventive replacement of wear parts) indicated by
the manufacturer.
1 Check accompanying documents:
Instructions for use are available
2 Check that the device combination is in good
condition:
All labels are complete and legible
No visible damage to the following parts:
Trolley and brakes
Housing parts
–Gas inlets
Hoses and cables
Fuses which are accessible from the
outside are in compliance with the specified
values
Country-specific labeling of gas types
3 Using the instructions for use, check that all
components and accessories needed to use
the product are available.
4 Check safety features:
Correct functioning of the emergency
breathing valve:
Before the check, switch off the device.
Pressure: max. –4 mbar (or hPa or cmH
2O)
Checks Interval Personnel responsible
Inspection and safety checks
1)
1) Designation applies to the Federal Republic of Germany; corresponds to the “Recurring safety inspection” in the Republic
of Austria
Every 6 months Service personnel
CAUTION
Risk of medical device failure
If safety checks are not performed on a regular
basis, the proper operation of the medical device
can be compromised.
Perform safety checks at the indicated intervals.
22609