Instructions for use Babylog 8000 plus SW 5.n 133
Technical data
Performance characteristics
Positive end-expiratory pressure or continuous
positive airway pressure
PEEP/CPAP
Range 0 to 25 mbar (0 to 25 cmH
2O)
Resolution 0 to 10 mbar (0 to 10 cmH
2O): 0.1 mbar
(0.1 cmH2O)
>10 mbar (>10 cmH
2O): 1 mbar (1 cmH2O)
Accuracy ±1 mbar (±1 cmH
2O) or 3 % of the measured value,
whichever is greater
Trigger sensitivity Trig
Range 1 to 10, corresponding to 0.02 to 3 mL
Composition of gas Breathing gas corresponding to the set O
2 concen-
tration, which can be set to NTPD or BTPS
Control principle Continuous flow, time controlled, pressure limited
O
2 is delivered by the integrated air/O2 mixer.
Trigger delay time 40 to 60 ms
Device compliance with breathing circuit without
breathing gas humidifier
Prerequisite:
– Inspiratory fraction: approx. 70 %
– Expiratory fraction: approx. 30 %
<1.0 mL/mbar (<1.0 mL/cmH
2O)
Device resistance with F&P 8411041 breathing cir-
cuit
Insp. Resistance 12 mbar (12 cmH
2O) at 30 L/min
Exp. Resistance 5 mbar (5 cmH
2O) at 30 L/min
Device resistance with F&P 8411153 HFV breathing
circuit
Insp. Resistance 6 mbar (6 cmH
2O) at 30 L/min
Exp. Resistance 4 mbar (4 cmH
2O) at 30 L/min
Device resistance with other breathing circuits:
Only use other breathing circuits if their resistance
values do not exceed the above-mentioned values
and the inspiratory fraction is approx. 70 %. Other-
wise the measurement of the airway pressure may
be impaired.