Getting started
56 Instructions for use Babylog 8000 plus SW 5.n
A self-test is carried out. All LEDs light up. A
continuous tone and an alarm tone sequence
sound briefly.
After the self-test is complete, the device
information is displayed, see page 99.
Then the main screen is displayed.
The device is ready for operation after
approx. 15 seconds. If the automatic calibration of
the O
2 sensor is activated, O2 measurement is
only available after 5 minutes.
Starting ventilation
1 Press the Vent. Mode key (A).
2 Press the IPPV/IMV key (B).
3 Press the On key.
4 Press the key.
The symbol for alarm limits (C) flashes.
Set the alarm limits, see page 84.
The Calibrate flow sensor! alarm message is
displayed.
Calibrate the flow sensor, see page 95.