Instructions for use Babylog 8000 plus SW 5.n 57
Getting started
Charging the battery for the power supply failure alarm
The battery is automatically charged when the
device is switched on.
Charge the battery for at least 30 minutes before
the first use and after being stored.
1 Make the following settings on the ventilator in
order to prevent alarms while the battery is
charging:
2 Switch the ventilator on.
3 Allow the charging time to expire.
WARNING
Absence of alarm during mains power supply
failure
If the battery is insufficiently charged, the
power supply failure alarm cannot sound in
the event of a mains power supply failure.
Charge the battery before the first use and
after being stored. Observe the charging time.
Parameter Setting
O
2-Vol.% 21
T
I 0.4 s
T
E 0.6 s
5L/min
Pinsp 20 mbar (20 cmH
2O)
PEEP/CPAP 3 mbar (3 cmH
2O)
Insp. Flow