Instructions for use Babylog 8000 plus SW 5.n 73
Operation
Performing pneumatic medication
nebulization
Conditions:
Medical gas supply from the central gas supply
system or compressed gas cylinders must be
ensured.
The kit for medication nebulization 8411025 is
available.
The medication nebulizer 8411030 has been
prepared in accordance with the corresponding
instructions for use.
The flow sensor must be removed from the
breathing circuit, see page 71.
Mounting the coupling
1 On the left side of the device, unscrew the
bottom mounting screw on the housing with a
coin.
2 Screw the coupling (A) into place with the
mounting screw.
3 Insert the plug (B) into the socket on the rear of
the device until it locks in place.
Installing the pneumatic medication nebulizer
1 Remove the corrugated hose of the breathing
circuit (C) from the inspiratory port of the Y-
piece and connect it to the inlet port of the
medication nebulizer.
2 Connect the corrugated hose (D), length
0.13 m (5.1 in), to the outlet port of the
medication nebulizer.
3 Connect the free end of the corrugated hose
(D) to the inspiratory port of the Y-piece.
When used on the incubator:
Push the inlet port or the outlet port of the
medication nebulizer into the upper hose guide
of the incubator.