Monitoring
94 Instructions for use Babylog 8000 plus SW 5.n
O2 monitoring
O2 monitoring is always activated on Babylog
8000. The function cannot be deactivated.
Calibrating the O2 sensor
Performing calibration
1 Press the Cal. Config. key.
2 Press the O
2-Cal key.
The device calibrates the O
2 sensor and displays a
message (A). O
2-Cal (B) is displayed in the status
field.
The calibration is completed after approx.
5 minutes. O
2-Cal (B) is no longer displayed in the
status field.
Hiding the message (A):
Press the OK key.
CAUTION
Faulty calibration
If the quality of the oxygen from the central gas
supply system is insufficient, the calibration may
be faulty.
Calibrate the O
2 sensor with calibration gas
(100 % O
2).