Alarms
68 Instructions for use Babytherm 8004/8010
Central alarm*
1 The red lamp on the radiant warmer flashes
when an audible alarm is emitted. As soon as
the alarm is acknowledged, the lamp goes out.
The central alarm always lights up in red for high-
priority and medium-priority alarms.
Alarm priorities
The color of the alarm LED and the alarm tone
sequence indicate the priority of the active alarm.
* Country-specific function (e.g., not available in China)
062
1
Color Signal Alarm tone
sequence
Priority of the alarm message Required action
Red Inop.LED is lit Continuous Device fault
High-priority
alarm
Immediate action is
necessary in order to
avert imminent danger
Red Power failure LED
is lit
Continuous Power
failure
Red Alarm LED flashes Intermittent Warning High-priority
alarm
!!! Immediate action is
necessary in order to
avert imminent danger.
Yellow Alarm LED flashes
(unacknowledged
alarm), alarm LED
is lit (acknowledged
alarm)
Intermittent
Caution
Medium-priority
alarm
!!
Prompt action is
necessary in order to
avert a danger.
--- --- --- Note Low-priority
alarm
Attention and action
are necessary.

Table of Contents

Other manuals for Dräger Babytherm 8010

Questions and Answers:

Dräger Babytherm 8010 Specifications

General IconGeneral
Control SystemMicroprocessor-controlled
ApplicationNeonatal care
Weight Capacity15 kg (patient weight)
Power Supply100 – 240 V AC, 50/60 Hz
Heat SourceCeramic
Control ModesManual Mode
AlarmsHigh Temperature, Low Temperature, Power Failure
Trendelenburg/Reverse TrendelenburgYes, adjustable
Height AdjustmentYes, electrically adjustable