Operation
101
Operating Instructions Caleo, Software 2.n
Selecting Menus
Ending trend analysis
To exit trend analysis
1 Press » « soft key to return to menu,
or
2 press key to terminate trend analysis.
Alternatively, if no key is pressed for 2 minutes, the display will
automatically revert to the standard screen.
Cleaning mode*
– Cleaning mode is only available if Caleo is equipped with
humidity control.
– Only use cleaning mode with an ambient humidity
of >10 % relative humidity.
– Close clamp on water supply connection kit before starting
cleaning mode.
– Cleaning mode lasts about 60 minutes.
– The water heater is heated up during cleaning mode. The
residue water evaporates. The water heater then cools
down.
– During cleaning mode Caleo
continues to measure the
relative humidity. The value will not be displayed. If the
relative humidity falls below 10 % the “Humidity sensor
inoperable” alarm will be triggered. Cleaning mode cannot
be ended in the correct manner. In order to maintain a
relative humidity of >10 %, the front and side doors and the
canopy should be open during cleaning mode.
– During or after cleaning mode, condensation could form
under the aggregate housing of the Caleo!
* Available option
WARNING!
Cleaning mode may only be used while Caleo is not
occupied by a patient.