Operation
101
Operating Instructions Caleo, Software 2.n
Selecting Menus
Ending trend analysis
To exit trend analysis
1 Press » « soft key to return to menu,
or
2 press key to terminate trend analysis.
Alternatively, if no key is pressed for 2 minutes, the display will
automatically revert to the standard screen.
Cleaning mode*
Cleaning mode is only available if Caleo is equipped with
humidity control.
Only use cleaning mode with an ambient humidity
of >10 % relative humidity.
Close clamp on water supply connection kit before starting
cleaning mode.
Cleaning mode lasts about 60 minutes.
The water heater is heated up during cleaning mode. The
residue water evaporates. The water heater then cools
down.
During cleaning mode Caleo
continues to measure the
relative humidity. The value will not be displayed. If the
relative humidity falls below 10 % the “Humidity sensor
inoperable” alarm will be triggered. Cleaning mode cannot
be ended in the correct manner. In order to maintain a
relative humidity of >10 %, the front and side doors and the
canopy should be open during cleaning mode.
During or after cleaning mode, condensation could form
under the aggregate housing of the Caleo!
* Available option
WARNING!
Cleaning mode may only be used while Caleo is not
occupied by a patient.