Care
119
Operating Instructions Caleo, Software 2.n
Testing of procedures and agents
The following materials are used in the patient capsule:
The following disinfectants have been tested and are not to be
used because of material incompatibility:
Components Material
Hood, flaps Polycarbonate
Pillar elements Styrene-butadiene thermo foam
injection molded material
ABS (acrylonitrile/butadiene/styrene)
Bed Styrene-butadiene thermo foam
injection molded material
ABS (acrylonitrile/butadiene/styrene)
Trough Styrene-butadiene thermo foam
injection molded material
ABS (acrylonitrile/butadiene/styrene)
X-ray drawer Polystyrene thermo foam injection
molded material
Housing Polystyrene thermo foam injection
molded material
Caleo
®
SoftBed
™
Polyurethane/polyester
Surface
disinfectant
Manufacturer
PaMo dur Dräger, Germany
PaMo dur rapid AF Dräger, Germany
PaMo dur Spray Dräger, Germany
PaMo sept Univ. AF. Dräger, Germany
Daisy Des Dräger, Germany
Incidin Extra N Henkel, Germany
Terralin Schülke & Mayr, Germany
Asphene Spray Laborat, France
Korsolin 50 Bode, France
Minudes Paragerm, France
Virufen Paragerm, France
Osuban S Japan
Taski Exact 300 Finland
WARNING !
Risk of burning when touching the heater!
When the incubator is closed, the heater is still hot
enough to cause serious burns for a long time after
switching off (still approx. 70 oC after 1 hour).
Allow Caleo® to cool down sufficiently before stripping it
down.
Manual cleaning and disinfection
Manual disinfection should preferably be carried out with
disinfectants on the basis of Aldehydes.
Observe the applicable country-specific listings for
disinfectants. The list of the German Association for Applied
Hygiene (Verbund für Angewandte Hygiene VAH) applies in
German-speaking countries.
The composition of the disinfectant is the responsibility of the
manufacturer and can change over time.
Strictly observe the manufacturer’s information on the
disinfectant.
Carry out manual cleaning and disinfection
z Remove soilings with a wet wipe.
1 Carry out surface disinfection.
2 After contact time, remove disinfectant residues.
Carry out machine cleaning and disinfection
1 Strictly observe Instructions for Use of washer-disinfector.
2 Position items so that all interior spaces are completely
flushed and water can drain off freely.
3 Use suitable cleaning agent.
4 Select suitable device program.
5 Carry out final rinsing with deionized water.
6 Immediately remove items from the washer-disinfector.
7 Inspect items for visible soiling and damage. If necessary,
repeat program or carry out manual cleaning and
disinfection.
8 Allow items to dry thoroughly.
Visual inspection
z Inspect all items for damage and wear, e.g. cracking,
embrittlement or pronounced hardening, and residual
soiling.
WARNING !
Penetrating liquid may cause malfunction of or damage
to the device, which may endanger the patient. Only
wipe-disinfect items and make sure no liquids penetrate
into the device.