Operating Instructions Caleo, Software 2.n
Before Reusing With a Patient
Assembling the double wall
z Check parts for damage and replace if necessary.
z Fit the parts into the double wall.
Fit in the following order:
1 Sleeve
2 Pin
3 Centring lugs
4 Seal
Insert the drainage module* into the pillar element
z Check the parts for damage and replace if necessary.
z Reassemble the drainage unit and drainage clip.
Make sure that they are correctly aligned.
5 Insert the drainage module into the pillar element from the
side until it engages.
The lug of the drainage clip must be facing outwards, and
the perimeter of the drainage module must be flush with the
pillar element.
z Close the side door.
z Check that the incubator is ready for operation –
Refer to the checks below and to "Before Each Use",
page 54.
* Available option
WARNING !
The incubator is ready for operation only when all
checks have been carried out successfully.