Troubleshooting
147
Operating Instructions Caleo, Software 2.n
Troubleshooting – Faults
Troubleshooting – Faults
All faults are listed in the table below in alphabetical order. See also "Alarm Description", page 161.
Fault Likely Causes Remedy
Bassinet cannot be set to the
tilted position.
Motor overheating. Wait for at least 2 minutes. After this
period, tilting is again possible for a few
seconds (10 %).
The full duty cycle is only available after
a pause of 54 minutes.
Motor faulty. Take unit out of service.
Height adjustment does not
switch off.
Switch faulty or loose contact.
Motor faulty.
Take unit out of service.
Height adjustment not possible. Lifting motor overheating. Wait for at least 2 minutes. After this
period, height adjustment is again
possible for a few seconds (10 %).
The full duty cycle is only available after
a pause of 54 minutes.
Lifting motor faulty. Take unit out of service.
Red bar LED on control unit is lit.
Continuous audible alarm tone.
Audible alarm cannot be silenced.
Serious device fault. Take unit out of service.
Red LED at symbol flashing,
intermittent alarm tone.
Power failure alarm. Check power supply.
Connect to line power.
Take unit out of service.
Short beep signal (3X) is sounded. While setting a set value, the rotary knob
is not pressed within 20 seconds.
Press rotary knob or cancel input.
The yellow LED in the alarm button
is lit.
Alarm suppression is active. Deactivate alarm suppression.
If any other faults not listed in this table occur, or in case Caleo does not respond as expected: Take unit out of service.