Technical Data
166
Operating Instructions Caleo, Software 2.n
Ambient Conditions
Operating Data
Technical Data
Ambient Conditions
Normal operation
Temperature 20
o
C to 35
o
C (59
o
F to 95
o
F)
Air pressure 600 hPa to 1060 hPa
Rel. humidity 10 to 95 % without condensation
Storage / transport
Temperature –20
o
C to 60
o
C (–4
o
F to 140
o
F)
Air pressure 210 hPa to 1060 hPa
Rel. humidity 10 to 95 % without condensation
Operating Data
Electrical power supply 100 V / 110 V / 120 V / 127 V / 220 V to 240 V
(please specify in your order) 50 Hz / 60 Hz
Max. current consumption at 120 V 8.7 A*
Leakage current 100 – 127 V / 50 Hz / 60 Hz 150 µA
Heater power
Line voltage connection 100 V/120 V 110 V/127 V 230 V 240 V
Air heater max.
525 W 588 W 525 W 572 W
Water heater max. 147 W 165 W 147 W 160 W
Built-in power socket strip
Max. permissible total power
consumption for all sockets
2 A
Max. leakage current:
By connecting devices to the power socket strip the overall leakage current may be increased to an unacceptably high level.
Where applicable, national limits must be observed. The operator is responsible for adhering to the specified maximum overall
leakage current.
USA (UL 60601-1):
Permissible overall ground leakage current 500 µA
Max. ground leakage current of the socket strip 350 µA
Height adjustment and tilting
Duty cycle 10 %
Shut-off mode 6 minutes ON, 54 minutes pause (intermittent duty)
O
2 inlet pressure min. 300 kPa (43.5 psi), max. 600 kPa (87 psi)
* Values include power consumption of the integrated power strip