Technical Data
169
Operating Instructions Caleo, Software 2.n
Physical Characteristics
Nurse call (optional) Output for connection to in-house P.A. systems (nurse call)
Connection 3 to 5 will be closed and the nurse call is activated in the
event of an alarm with a high risk potential or if there is a serious
equipment fault.
Connection 1 to 3 will be closed and the nurse call is not activated in
the event of an alarm with a medium or low risk potential.
Potential-free changeover contact
243
Central alarm (optional) Output for connection to in-house P.A. systems (nurse call)
Operating voltage max. 24 V
Current max. 250 mA
Power max. 3 W
Physical Characteristics
Dimensions
Incubator (width x depth) 1167 mm x 687 mm (45.9" x 27.0")
Height overall with height adjustable column 1220 mm to 1520 mm (48.0" x 59.8")
Height overall with fixed column 1270 mm / 1370 mm / 1470 mm (50" / 53.9" / 57.9")
Height of mattress surface (variable height) 800 mm (31.5") to 1100 mm (43.3")
Height of mattress surface (fixed height) 875 mm (34.4") / 950 mm (37.4") / 1050 mm (41.3"),
depending on fixed height column installed
Bed (width x depth) 645 mm x 500 mm (25.4" x 19.7")
Weight
Overall weight
max. 230 kg (507 lbs)
basic configuration 130 kg (286.6 lbs)
Overall load capacity 66 kg (142.5 lbs)