For detailed information, always refer
to Caleo Operating Manual !
Before use, check device according to
Instructions in Operating Manual.
Switch on device (rear side of mobile
stand).
Select control mode by pressing button.
Adjust settings with rotary dial knob.
Confirm settings by pressing dial knob.
See text messages in case of alarm.
DANGER!
Risk of explosion if used in the presence of flammable
anesthetics.
WARNING - FIRE HAZARD!
Keep all sources of ignition out of the room in which the
incubator is located.
Higher risk of ignition and fire in air enriched with
oxygen.
All oxygen valves, connections and seals must be kept
free of oil and grease.
Open valves slowly.
Do not use any electrical equipment inside the
incubator other than equipment or instruments
expressly designed and approved for use inside
incubators.
WARNING!
Disconnect supply before servicing.
Repairs on this equipment to be performed by
DraegerService or factory trained and authorized
personnel only.
CAUTION!
Opening of covers by DraegerService or factory trained
and authorized personnel only.
To maintain grounding integrity connect only a 'hospital
grade' power outlet.
Federal (US) Law restricts this device to sale by or on
the order of a physician.
WARNING!
Danger of patient injury: never leave infant unattended
when doors or handports are open.
WARNING!
Check infant temperature and skin condition at regular
intervals and adjust temperature setting to individual
patient needs.
Check incubator temperature, incubator function, and
temperature sensor attachment at regular intervals.
Improper attachment or location of the temperature
sensor may cause cooling or overheating of the patient.
WARNING!
External radiant heat sources such as radiant heaters
or sunlight may increase incubator air temperature
above the set level
WARNING!
Never block or obstruct air vents! Risk of burning!
Dräger Medical, Inc.
3135 Quarry Road
Telford, PA 18969
Caleo Infant Incubator
Air
Skin
Quick Reference