Operation
74
Operating Instructions Caleo, Software 2.n
Switching Between Air and Skin Temperature Control
After the new mode has been activated, you can set the
desired value with the rotary knob.
1 The LED of the activated mode will start blinking.
2 Turn rotary knob to adjust setpoint.
2 Press rotary knob to confirm the new setpoint.
– The upper part of the screen displays the advisory
message:
»confirm new mode with rotary knob«.
2 Press rotary knob to confirm the new mode.
– The display returns to the standard screen.
1 The LED of the activated mode is now continuously lit.
If you do not wish to change the settings:
The display returns to the standard screen. The previous
setpoint is retained.
Or
– Wait for 7 seconds: Caleo emits 4 short beeps to prompt the
user to press the rotary knob. The display will immediately
return to the standard screen. The previous setpoint will be
retained.
z Press and the new settings will be cancelled.