C.O. SETUP - HARDWARE
VF8 DELTA/DELTA XL/KAPPA 15-9
If you choose Other as the catheter type, you must enter a computation constant in
order to display or select the catheter size and injectate volume. Consult the
documentation included with the catheter for computation constants, and select one
that corresponds to the injectate volume and temperature being used.
5F 5 cc 0.291 0.316 0.279
5F 3 cc 0.170 0.188 0.160
Arrow
Injectate Temperature (IT) Sensor connected
Catheter Size
Injectate
Volume IT = -1 °C (
±1
°C) IT = 24 °C (
±1
°C)
7.5F 10 cc 0.532 0.586
7.5F 5 cc 0.249 0.265
7.5F 3 cc 0.131 0.155
7F 10 cc 0.541 0.601
7F 5 cc 0.250 0.273
7F 3 cc 0.134 0.156
5F 5 cc 0.267 0.303
5F 3 cc 0.157 0.192
BD/Ohmeda
Injectate Temperature (IT)
Sensor connected
IT Sensor
disconnected
Catheter
Size
Injectate
Volume IT = -5 °C to +16 °C IT = 16 °C to 27 °C IT = 0 °C