STATUS M ESSAGES
VF8 DELTA/DELTA XL/KAPPA 17-17
SpO
2
Regulation
Error
• Inconsistent light level
• Excess ambient light
• Make sure the SpO
2
sensor is properly
attached to the patient.
• Remove or cover any external sources
of light entering the sensor.
• Contact DrägerService.
SpO
2
Unplugged The sensor cable is not
connected to the MultiMed pod.
• Make sure the cables are securely
connected.
• Check for a defective sensor.
SpO
2
: non-Masimo
compatible sensor
A non-Masimo compatible
sensor or SpO
2
pod is
connected to a monitor that is
configured for Masimo.
• Replace with a Masimo-compatible
sensor.
• Contact the biomedical personnel
(SpO
2
compatibility is a locked option).
Duplicate Device
Connected
A MultiMed pod (with SpO
2
sensor) and an Infinity Masimo
SET pod are connected
simultaneously.
Disconnect the duplicate device.
Message Possible Cause Suggested Action