17-18 DELTA/DELTA XL/KAPPA VF8
17 P
ULSE OXIMETRY (SPO2)
Nellcor OxiMax pod
Message Possible Cause Suggested Action
SpO
2
> #
SpO
2
< #
The patient’s SpO
2
is outside the
current upper or lower alarm limits.
Check the patient and treat, if
necessary.
PLS > #
PLS < #
The patient’s pulse rate is outside
the current upper or lower alarm
limits.
SpO
2
< Low limit
value - 20%
(Neonatal patient
category only)
The patient’s SpO
2
is below the
current lower alarm limit by 20% or
more.
PLS Out of Range
(Low/High)
The pulse rate is outside the
measuring range of the monitor.
SpO
2
Unrecognized
Sensor
The monitor does not recognize the
sensor connected as valid.
• Check for a defective or
unapproved sensor.
• Replace the sensor.
• Disconnect/reconnect the pod and
replace with a good sensor.
• Contact the hospital’s technical
personnel or DrägerService.
SpO
2
Artifact/
Sensor/Site?
• Persistent artifact
• Wrong sensor and/or site was
used.
• High pulse amplitude detected
• Sensor improperly placed.
• Electrical/optical interference
• Light is blocked
• Make sure the SpO
2
sensor is of a
correct type and is properly
attached to the patient and that the
patient is not moving, and all cables
are properly connected.
• Make sure no electrical or optical
interference is present.
• Check the sensor for blocked light
(may be due to nail polish).
• Change the site/location
periodically (for example finger vs.
forehead, etc.).
• Contact DrägerService.
SpO
2
Weak signal/
Sensor Off?
• Pulse amplitude is too low.
• Weak signal/weak pulse due to one
of the following conditions.
- Poor perfusion.
- Low body temperature.
- Sensor off the patient.
• Check the patient.
• Make sure the SpO
2
sensor is
properly attached to the patient and
all cables are properly connected.
• Relocate the sensor to another
extremity.
• Contact DrägerService.