Instructions for use
Movita
Ceiling supply units
WARNING
To properly use this medical device,
read and comply with these
instructions for use.

Questions and Answers

  • M
    Matthew ReyesAug 15, 2025
    Why Dräger Movita Medical Equipment lift system does not respond when actuated?
    • J
      Jason SchwartzAug 15, 2025
      If the lift system of your Dräger Medical Equipment isn't responding, the system may be without power or the cable connection may be interrupted. You should check the power supply and fuses (hospital maintenance).
  • M
    Matthew HendrixAug 18, 2025
    What to do if Dräger Movita Medical Equipment has insufficient gas supply and a hissing sound?
    • K
      Kevin ArmstrongAug 18, 2025
      If you hear a hissing sound and suspect an insufficient gas supply in your Dräger Medical Equipment, it may be that pressurized gas is escaping from inside the ceiling supply unit. Notify hospital maintenance or DrägerService to check the gas supply, considering also connected devices.
  • D
    Donna HenryAug 21, 2025
    What to do if the pneumatic brake does not release on Dräger Medical Equipment?
    • O
      Oscar FryAug 21, 2025
      If the pneumatic brake on your Dräger Medical Equipment isn't releasing, it could be due to a non-working pneumatic valve or brake key (no hissing sound when actuated). Check the compressed air system. If the compressed air system is in order, inform DrägerService.
  • R
    Ryan CooperAug 25, 2025
    Why is the LED indicator not lit green on Dräger Movita lift system?
    • L
      Laura FuentesAug 25, 2025
      If the LED indicator on your Dräger Medical Equipment lift system is not lit green, the board/system may be without power, or the cable connection may be interrupted. Check the power supply and fuses (hospital maintenance).
  • M
    Mrs. Jasmine Stone PhDAug 27, 2025
    What to do if the electromagnetic brake does not release on Dräger Movita?
    • P
      Paul MontesAug 28, 2025
      If the electromagnetic brake on your Dräger Medical Equipment does not release, it could be that the electromagnet or key is not working. Inform DrägerService.
  • J
    Jennifer WeaverAug 30, 2025
    Why electromagnetic brake does not engage on Dräger Movita?
    • C
      Cristian TurnerAug 30, 2025
      If the electromagnetic brake on your Dräger Medical Equipment does not engage, it may be that the electromagnet is not supplied with current. Inform DrägerService.
  • J
    johnjonesSep 2, 2025
    Why Dräger Movita Medical Equipment makes a continuous hissing sound?
    • A
      Angela HoltSep 2, 2025
      If your Dräger Medical Equipment is making a continuous hissing sound, it may be due to defective pneumatic connections. Inform DrägerService.
  • L
    Lisa CarterSep 5, 2025
    What to do if Dräger Movita Medical Equipment lift system does not respond to the down arrow key?
    • P
      Peggy TranSep 5, 2025
      If the lift system on your Dräger Medical Equipment does not respond when you press the down arrow key, it's possible that the Perseus safety sensor has been triggered. Remove any obstacle and then try operating the up arrow key.
  • D
    Daniel MooreSep 8, 2025
    What to do if bearing with friction brake turns too easily or stiffly on Dräger Movita?
    • J
      Joseph JonesSep 8, 2025
      If the bearing with friction brake on your Dräger Medical Equipment turns too easily or stiffly, it may be that the friction brake is adjusted too loose or tight. Inform DrägerService.
  • S
    Sharon PierceSep 12, 2025
    What to do if bearing with pneumatic brake or electromagnetic brake turns too stiffly on Dräger Movita Medical Equipment?
    • N
      Nathan SpenceSep 12, 2025
      If the bearing with pneumatic brake or electromagnetic brake on your Dräger Medical Equipment turns too stiffly, it may be that the friction brake is adjusted too loose or tight. Inform DrägerService.

Summary

Safety Information Definitions

User and Patient Safety

General Safety Information

Essential safety instructions for general operation, including intended use and service.

Critical Safety Warnings and Risks

Highlights critical warnings regarding personal injury, equipment damage, and specific module installation.

Application of the Movita Unit

Assembly and Preparation Procedures

Device Positioning and Connections

Provides warnings and instructions for safe positioning and connection of devices.

Electrical, Gas, and AGSS Connections

Instructions for connecting power, medical gas probes, and AGSS/AIR motor probes.

Operating the Movita Ceiling Supply Unit

General Operation Safety and Panel Blocking

Covers touch voltages, personal injury risks, fire hazards, and control panel blocking.

Brake Release and Arm Positioning

Instructions for releasing brakes and positioning arm systems.

Single and Double Arm Positioning

Guides for positioning single-arm and double-arm units, including light integration.

Movita Lift Height Adjustment

Moving Media Column Up and Down

Step-by-step guide for moving the media column/head upwards and downwards.

Movita Lift Device Mounting and Cart Docking

MOVA Cart Docking and Undocking

Step-by-step guides for docking and undocking MOVA lift equipment carts.

Troubleshooting Guide

Movita Lift Emergency Operation

Instructions for manual system operation during power failure.

Cleaning, Disinfection, and Sterilization

Maintenance Schedule and Component Care

Disposal of Medical Devices

Maximum Load Capacities

Load Limits for GH3;GH3+ Lifting Modules

Specifies maximum load on support arms with mounting rails for GH3/GH3+ lifting modules.

Movita Arm System Load Limits

Table of maximum loads for Movita arm systems (light, medium, strong) by length.

Dräger Movita Specifications

General IconGeneral
BrandDräger
ModelMovita
CategoryMedical Equipment
LanguageEnglish