EasyManua.ls Logo

Dräger MIcroVent - User Manual

Dräger MIcroVent
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Operating Instructions MicroVent
MicroVent
Ventilator
Operating Instructions
Software 1.n
2-680-95
D

Table of Contents

Questions and Answers

  • J
    josepholsonSep 23, 2025
    What to do if Dräger Medical Equipment displays 'Ventilator INOP' with continuous alarm?
    • M
      mwarnerSep 23, 2025
      If your Dräger Medical Equipment displays 'Ventilator INOP,' ventilation stops, the alarm light is continuously lit, and there's a continuous audible alarm, it indicates an internal ventilator malfunction. Switch off the ventilator and disconnect it from the external power supply. Switch the ventilator on again. If the fault recurs, use a backup ventilator and call DrägerService.
  • L
    Lisa CarterAug 25, 2025
    What to do if the battery is discharged on Dräger MIcroVent?
    • E
      Erica GonzalezAug 25, 2025
      If the battery is discharged and there's no external power supply for your Dräger Medical Equipment, replace the battery pack with a new one or connect to an external power supply.
  • A
    Aaron FloresAug 25, 2025
    What to do if Dräger MIcroVent Medical Equipment displays 'Change bat. Alarm'?
    • A
      adownsAug 26, 2025
      If your Dräger Medical Equipment displays a 'Change bat. Alarm', replace the battery pack with a new one or connect the equipment to an external power supply, as the battery pack will be discharged within a few minutes and no external power is connected.
  • L
    Lisa FreemanAug 27, 2025
    What to do if Dräger MIcroVent Medical Equipment shows 'NiCd discharged' and ventilation stops?
    • S
      Susan GarrisonAug 27, 2025
      If your Dräger Medical Equipment displays 'NiCd discharged' and ventilation stops, replace the NiCd pack with a new battery pack or connect the equipment to an external power supply, as the NiCd pack is discharged and no external power supply is available.
  • H
    hnelsonAug 27, 2025
    What causes 'Supply press.low' alarm on Dräger MIcroVent?
    • Z
      zlopezAug 27, 2025
      The 'Supply press.low' alarm on Dräger Medical Equipment indicates that ventilation has stopped. This can be due to: * An empty O2 cylinder or closed cylinder valve: Check the gas supply in the O2 cylinder and connect to a full O2 cylinder. Open the cylinder valve. * A gas probe not properly connected to the medical gas pipeline system: Check the gas pressure in the medical gas pipeline system to ensure it is greater than 2.7 bar. Fully insert the gas probe into the terminal unit.
  • K
    Karen HoodSep 15, 2025
    What causes the audible alarm of the O2-Air mixer on Dräger MIcroVent Medical Equipment?
    • M
      Megan JonesSep 15, 2025
      The audible alarm of the O2-Air mixer (for blender 84 12 411) on Dräger Medical Equipment can be caused by a supply pressure of O2 or air being too low, so ensure adequate supply pressures. It may also be due to a difference in pressure between the two supply pressures being greater than 14 psi, so equalize supply pressures so that the difference is less than 14 psi.
  • H
    Heather AyalaSep 12, 2025
    What to do if Dräger MIcroVent Medical Equipment displays 'Flow measurement INOP'?
    • C
      Colleen AguilarSep 12, 2025
      If your Dräger Medical Equipment displays 'Flow measurement INOP', it means flow measurement is not operable and signals from flow measurement are not used for alarms. Ventilation may be continued. After shutdown, call DrägerService.
  • J
    Jessica HardinSep 14, 2025
    What does 'No NiCd/bat.' mean on Dräger Medical Equipment?
    • S
      smithdavidSep 14, 2025
      The message 'No NiCd/bat.' on Dräger Medical Equipment means the machine is connected to an external voltage supply, but no internal power supply is available. Install a fully-charged NiCd pack or a new alkaline battery pack.
  • B
    Brian BenitezSep 5, 2025
    What does 'Check settings Alarm' mean on Dräger MIcroVent Medical Equipment?
    • M
      Matthew RileySep 5, 2025
      The 'Check settings Alarm' on Dräger Medical Equipment means the settings are outside ventilator performance limits, for example, the required flow is less than 10 L/min or greater than 80 L/min. Adjust set parameters, such as f, VT, or Tinsp.
  • T
    Troy CamachoSep 2, 2025
    What causes 'MV low Alarm' on Dräger MIcroVent Medical Equipment?
    • S
      Sandra HernandezSep 2, 2025
      A 'MV low Alarm' on Dräger Medical Equipment can be triggered due to disconnection/leakage, so make sure connections are leak free. It may also be caused by a defective or dirty flow sensor, so exchange the flow sensor. Another cause can be insufficient spontaneous breathing of patient during CPAP, so switch to SIMV or CMV.

Dräger MIcroVent Specifications

General IconGeneral
BrandDräger
ModelMIcroVent
CategoryMedical Equipment
LanguageEnglish

Related product manuals