Operating Instructions MicroVent
MicroVent
Ventilator
Operating Instructions
Software 1.n
2-680-95
D

Table of Contents

Questions and Answers

  • T
    Tyler MorrisSep 9, 2025
    What causes the 'Press.Supp.> 4 s' alarm on Dräger MIcroVent Medical Equipment?
    • R
      Robin HerringSep 9, 2025
      The 'Press.Supp.> 4 s' alarm on Dräger Medical Equipment can be caused by a leak in the patient connection or hose system (make sure connections are tight), a diaphragm in the ventilation valve installed incorrectly or damaged (install the diaphragm correctly or replace it), or a leaking cuff (inflate the cuff and test for leaks).
  • M
    melissaesparzaSep 7, 2025
    What triggers 'Paw low Alarm' on Dräger MIcroVent Medical Equipment?
    • A
      Angelica ChavezSep 8, 2025
      A 'Paw low Alarm' on Dräger Medical Equipment is triggered by disconnection or leakage at the patient connection, ventilation valve, or ventilator circuit. Ensure all connections are leak-free.
  • K
    Kathleen CarpenterAug 22, 2025
    How to fix AC power failure in Dräger MIcroVent?
    • X
      xvegaAug 22, 2025
      If there is an AC power failure with Dräger Medical Equipment, connect it to an electrical power supply as soon as possible.
  • B
    Brian BenitezSep 5, 2025
    What does 'Check settings Alarm' mean on Dräger MIcroVent Medical Equipment?
    • M
      Matthew RileySep 5, 2025
      The 'Check settings Alarm' on Dräger Medical Equipment means the settings are outside ventilator performance limits, for example, the required flow is less than 10 L/min or greater than 80 L/min. Adjust set parameters, such as f, VT, or Tinsp.
  • K
    kelly87Sep 4, 2025
    What to do for an Apnea Alarm on Dräger MIcroVent?
    • R
      Randy CoxSep 4, 2025
      If you're getting an Apnea Alarm on Dräger Medical Equipment, it means the patient's spontaneous breathing in CPAP mode is insufficient. Switch to CMV or SIMV.
  • T
    Troy CamachoSep 2, 2025
    What causes 'MV low Alarm' on Dräger MIcroVent Medical Equipment?
    • S
      Sandra HernandezSep 2, 2025
      A 'MV low Alarm' on Dräger Medical Equipment can be triggered due to disconnection/leakage, so make sure connections are leak free. It may also be caused by a defective or dirty flow sensor, so exchange the flow sensor. Another cause can be insufficient spontaneous breathing of patient during CPAP, so switch to SIMV or CMV.
  • C
    Charles CoxSep 1, 2025
    What does 'Leakage Alarm' mean on Dräger MIcroVent?
    • A
      Adam WilliamsSep 1, 2025
      The 'Leakage Alarm' on Dräger Medical Equipment indicates that the measured expiratory minute volume has fallen to about 40% below the inspiratory minute volume. Correct any leakage in the patient system and/or in the ET-tube, if necessary.
  • S
    skinnerjamesAug 31, 2025
    What causes 'MV high Alarm' on Dräger Medical Equipment?
    • S
      Steven PaulAug 31, 2025
      A 'MV high Alarm' on Dräger Medical Equipment can be triggered if the patient hyperventilates or the upper alarm limit is too low, so adjust the upper alarm limit. Another possible cause is a defective or dirty flow sensor, so exchange the flow sensor.
  • R
    Robert JohnsonAug 30, 2025
    What causes 'Paw high Alarm' on Dräger Medical Equipment?
    • T
      Tony MillerAug 30, 2025
      A 'Paw high Alarm' on Dräger Medical Equipment, where full minute volume cannot be applied, can be caused by: * Stenosis in airways, a kinked patient circuit, reduced lung compliance, or inspiratory flow from settings of VT, f, and Tinsp being too high: Clear the patient's airways, unkink the patient circuit, set Pmax higher, and use the Tinsp knob to set a longer inspiratory time to reduce flow. * The patient "fighting" the ventilator: Change the ventilation pattern or mode of ventilation, and sedate the patient if necessary.
  • S
    Scott AdamsAug 28, 2025
    What does 'Charge NiCd Alarm' mean on Dräger MIcroVent?
    • D
      Daniel WoodAug 28, 2025
      The 'Charge NiCd Alarm' on Dräger Medical Equipment means the NiCd pack will be discharged within a few minutes, and no external power supply is connected. Replace the rechargeable battery pack with a fully charged NiCd pack, or connect to an external power supply.

Dräger MIcroVent Specifications

General IconGeneral
BrandDräger
ModelMIcroVent
CategoryMedical Equipment
LanguageEnglish