Testing Pressure Support Ventilation
1
Set switch for ventilator modes to SIMV/CPAP,
2 knob »f« to 0 bpm,
3 knob »PS« to 25 cmH2O,
4 Knob «PEEP« to 10 cmH2O,
● Squeeze test lung a little, release and observe
diameter:
The airway pressure should rise to 25 cmH2O
±3 cmH2O, (pressure support) and then drop to
10 cmH2O ±2 cmH2O (PEEP).
● On / Off switch to 0 (OFF).
● Press g key to confirm.
● Remove test lung, re-attach elbow connector.
WARNING !
Ventilator is ready for operation only after all
tests were performed successfully.
Testing Readiness for
Operation
Testing Pressure Support
Ventilation
29