EasyManua.ls Logo

Dräger MIcroVent - Testing Ventilator Function

Dräger MIcroVent
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Testing Ventilator Function
Set rotary knobs:
1 knob »VT« to 0.5 L.
2 knob »f« to 12 bpm;
3 knob »Tinsp« to 1.5 seconds.
4 knob »Pmax« to 60 cmH2O.
5 knob »PEEP« to 5 cmH2O.
6 Set switch for ventilator modes to »CMV«.
7 Switch ventilator »ON«.
8 Set alarm limits "MV" to 4 and 8 L/min.
The ventilator will now perform an electric and
pneumatic self test.
Display:
The bottom line displays the software version cur-
rently in use.
The red alarm status light and the LCD backlighting
will be turned on briefly.
The audible alarm will sound twice
NOTE: For safety reasons, the audible alarm output
uses two channels which are both tested during the
self test. Therefore two audible alarms of the same
duration will be heard.
The green status light » « will be on continuously
as long as an external power source is connected.
The self test will be completed after no more than
6 seconds
Display:
MicroVent will now ventilate the test lung with
the preset ventilation pattern.
The pressure gauge should alternate between a
consistent inspiratory pressure and an endexpiratory
pressure of about 5 cmH2O.
The display should show a minute volume of:
(Tolerance: ±1.0 L/min)
If the value is outside the tolerance:
Replace flow sensor
CMV MV = 6.0
Self test O.K.
Self test
Software xx.xx
Testing Readiness for
Operation
Testing Ventilator Function
25
Operating Instructions MicroVent
D
2 1 3 6
4 8 5 7

Table of Contents

Related product manuals