Displaying Measured Ventilation
Parameters
Both ventilation mode set and expiratory minute volume
(MV) are displayed continuously in the top line of the
LCD screen.
In the bottom line, different ventilation parameters can be
displayed, depending on the current mode of ventilation
(see the following examples):
Switching between measured parameter values to be
displayed:
Briefly press l key.
During CMV
Inspiratory Flow FLOW = 30.0 L/min
I:E-ratio TI:TE = 1:2
Peak pressure PEAK = 20 cmH2O
Mean pressure MEAN = 15 cmH2O
Endexp. pressure PEEP = 5 cmH2O
Breathing rate Frequency = 10 bpm
During SIMV
Inspiratory Flow
FLOW = 30.0 L/min
Peak pressure PEAK = 20 cmH2O
Mean pressure MEAN = 15 cmH2O
Endexp. pressure PEEP = 5 cmH2O
Breathing rate Frequency = 10 bpm
Tital volume VT = 450 ml
During CPAP
Positive airway pressure CPAP = 8 cmH2O
Peak pressure PEAK = 20 cmH2O
Mean pressure MEAN = 15 cmH2O
Breathing rate Frequency = 10 bpm
Tidal volume VT exp. = 450 mL
Operation Displaying Measured
Ventilation Parameters
36
Operating Instructions MicroVent