Dräger MIcroVent - Technical Data; Ambient Conditions; Performance Data

Dräger MIcroVent
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Expiratory minute min. 2.0 L/min
ventilation MV max. 25.0 L/min
with I:E = 1:1.5 and effective
flow range of 10 to 80 L/min
Characteristics for support of spontaneous breathing
Demand valve
trigger pressure approximately –1 cmH2O
Max. demand flow supply
at –4 cmH2O 120 L/min
Synchronization sensitivity
for A/C 5 L/min
Base flow for spontaneous
breathing approximately 5 L/min
Ventilator compliance
(incl. patient circuit) <3 mL/cmH2O
Inspiratory resistance <6 cmH2O/L/s
Expiratory resistance <6 cmH2O/L/s
Dead space including
flow sensor approximately 28 mL
Measuring ranges
Pressure gauge –10 to 80 cmH2O
Accuracy ±2 cmH2O
Flow measurement range 2 to 120 L/min bidirectional
Max. differential
pressure allowed for
flow measurement ±4 cmH2O
Resistance of flow sensor 3 cmH2O at 100 L/min
Measurement of minute
ventilation
Range 2 to 40 L/min
Accuracy (at 100 % O2
1013 hPa, 68 ºF,
50 %r.H.)
for 2 to 5 L/min ±1 L/min
for 5 to 40 L/min greater of ±12 % of mea-
sured value or ±1 L/min
Patient connector male ISO taper, 22 mm
Technical Data
NOTE: Specifications are given conforming to ISO 5369.
Ambient Conditions
For operation
Temperature
Ventilator 14 to 122 ºF (–10 to 50 °C)
AC-adapter 32 to 104 ºF (0 to 40 °C)
Atmospheric pressure 600 to 1200 hPa
Rel. humidity 30 to 95 % r.H.
For storage
Temperature 0 to 158 ºF (–20 to 70 °C)
AC-adapter –40 to 158 ºF (–70 to 40 °C)
Atmospheric pressure 600 to 1200 hPa
Rel. humidity 10 to 95 % r.H.
Performance Data
Operating parameters
Control principle volume flow controlled, time-
cycled, volume constant,
flow interrupter
(microprocessor controlled)
spontaneous breathing with
integrated demand valve
(also at elevated PEEP levels)
Ventilation modes CMV/AC, SIMV/CPAP
Ventilator rate
CMV 5 to 60 bpm ±1 bpm
SIMV 2 to 60 bpm ±1 bpm
continuously adjustable
Tidal volume VT 0.1 to 1.5 L
continuously adjustable
±10 % of set value,
50 mL minimum
Inspiratory time Tinsp. 0.2 to 3 seconds ±5 %
continuously adjustable
inspiratory pressure
limitation Pmax 20 to 80 cmH2O ±10 %
minimum ±3 cmH2O
continuously adjustable
PEEP 0 to 18 cmH2O
CPAP 3 to 18 cmH2O
Pressure Support 0 to 35 cmH2O
Technical Data Ambient Conditions
Performance Data
56
Operating Instructions MicroVent