Electrical Supply
The MicroVent ventilator is designed for different
types of electrical supply.
Internal supply
with either rechargeable (NiCd) or non-rechargeable
(alkaline) internal battery pack.
External supply
with AC adapter (part no. 84 12 709) from 120 V AC
line voltage for recharging of battery pack and for ex-
tending operation when using battery packs.
Use with internal rechargeable battery pack or alka-
line batteries
CAUTION !
For safety reasons, a battery pack must always be
installed, even when using external power supply!
CAUTION !
Always use AC adapter 84 12 709 for connecting
device to 120 V AC line voltage.
Electrical Supply
Preparation
17
Operating Instructions MicroVent
CAUTION !
The internal battery pack will last for up to 2 hours
of operation. Keep additional, fully charges battery
packs on hand for backup if external sources might
not be available for more than 2 hours
CAUTION !
When exchanging the internal battery pack during
operation, always assure that ventilation can be
provided with an independent ventilation device
(resuscitation bag)while the MicroVent is discon-
nected from power.