EasyManua.ls Logo

Dräger MIcroVent - Gas Supply

Dräger MIcroVent
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Gas Supply
Gas supply with O2-Air mixer
1 Connect air and oxygen high pressure hoses to
oxygen blender.
2 Connect blender outlet to gas inlet on MicroVent.
WARNING !
Do not install flow restricting valves or flow me-
ters into the high pressure gas supply. Equip-
ment malfunction might result.
WARNING !
Always use medical grade oxygen that is dry and
free from dust and oil. Contaminated gas may
cause ventilator malfunction.
WARNING !
Always use extreme caution when using oxygen!
Oxygen intensely supports any burning!
No smoking, no open fire in areas where oxygen
is in use!
Always provide adequate ventilation in order to
maintain ambient O2 concentrations < 24 %
Always secure O2 cylinders against tipping, do
not expose to extreme heat.
Do not use oil or grease on O2 equipment such as
tank valves or pressure regulators. Do not touch
with oily hands. Risk of fire!
Open and close valves slowly, with smooth turns.
Do not use any tools.
PreparationGas Supply
21
Operating Instructions MicroVent
D
1
Air O2
2

Table of Contents