Operation
Controlled Mandatory Ventilation (CMV)
For ventilator rates of 5 to 60 bpm.
NOTE: During CMV, the MicroVent ventilator limits the
minimum ventilator rate to 5 bpm.
1 Set switch for ventilator modes to CMV
2 Set knobs »f« and »VT«.
3 Set knob »Tinsp.«.
4 Set knob »Pmax« initially to 60 cmH2O.
5 Set knob »PEEP« initially to 0 cmH2O.
6 Set O2 concentration on blender.
7 Set alarm limits for minute volume.
8 On/Off switch to »ON«.
Display:
When the patient has been connected:
Check display of minute ventilation and re-adjust
for the specific patient.
WARNING !
Observe pressure gauge in order to identify im-
proper ventilation. Operator must always be
ready to intervene if alarm indicates a situation
dangerous to the patient.
Self-test
Software xx.xx
WARNING !
Always use ventilator that has been cleaned
and disinfected and has been successfully
tested to be ready for operation.
Controlled Mandatory
Ventilation (CMV)
Operation
30
Operating Instructions MicroVent