Continuous audible alarm and flashing of red alarm
status light both continue.
When spontaneous breathing is insufficient:
switch to SIMV or CMV.
To switch off apnea ventilation:
press g key twice.
Pressure-limited ventilation
Use optional pressure limiting valve 8405390!
Draw ventilation hose off twin nozzle of the
humidifier.
Draw elbow connector (with twin nozzle) off the
humidifier connection.
1 Plug twin nozzle (with elbow connector) into the
humidifier connection.
2 Plug pressure limiting valve onto elbow connector
with adjusting screw pointing upwards.
3 Plug ventilation hose onto nozzle of pressure limiting
valve.
Set ventilation as described for IPPV/SIMV mode.
Set inspiratory pressure by means of adjusting screw
on pressure limiting valve and check on pressure
gauge.
Check set tidal volume VT.
Correct alarm limits for the minute volume MV and
max. airway pressure Paw if necessary.
OperationPressure-limited ventilation
35
Operating Instructions MicroVent
2
1
3