Customer Service Mode
Customer Service Mode
– for selecting the language for display texts
– for checking the machine status and machine
functions
– for switching ON / OFF the high rate-Alarm
The information provided by these checks allows
DrägerService to draw conclusions as to the cause of a
fault.
The display texts for customer service mode appear in
English.
Select Customer Service Mode
Disconnect MicroVent from patient.
1 On / off switch to 0 ( OFF).
2 "f" rotary knob to max. = 60 / min.
3 "VT" rotary knob to max. = 1.5 L
4 Press l and g keys and hold down
While holding down these keys:
1 On / off switch to I (ON)
MicroVent performs self-test, then display:
● Release keys.
Display:
Activate
4 Confirm = briefly press l and g keys
simultaneously.
Customer
Service Mode
Adjust language
<- 001 +>
D
32
4 4
1
Customer Service Mode
47
Operating Instructions MicroVent