O2 Supply from an oxygen cylinder
Always use full tank (2400 psi)
Attach pressure regulator for delivery pressure of 38
to 84 psi ( = 2.7 to 6.0 bar), nominal pressure 50 psi
to oxygen cylinder.
Connect MicroVent with DISS high pressure hose to
oxygen regulator.
Open valve slowly and completely.
Determining gas supply in O2 cylinder
Example:
Tank pressure according to pressure gauge
on regulator:
Supply of pressurized O2 (E-cylinder):
Estimate for expected time of operation of
MicroVent
Example:
Mode CMV:
Minute Ventilation:
expiratory gas consumption due to the base flow
pressur. gas supply [L]
Expected time of op. =
(MV + 1)
+
1 –
Tinsp. · f
· 5[L/min]
60
Close cylinder valve completely after test
Supply with oxygen from wall pipeline system
Attach high pressure hose with DISS connector to
MicroVent and connect to wall terminal with
appropriate quick connect probe.
_______
* Internal ventilator gas consumption approximately 1 L/min
Gas Supply
Preparation
22
Operating Instructions MicroVent
Rate = 10/min, VT = 1 L, Tinsp. = 3 s
MV = 10/min x 1 L = 10 L/min
= 1 –
Tinsp. · f
· 5 [L/min]
60
Top = 622 / 13,5 = 46 minutes
2400 psi (165 bar)
622 L