What causes 'Supply press.low' alarm on Dräger MIcroVent?
- ZzlopezAug 27, 2025
The 'Supply press.low' alarm on Dräger Medical Equipment indicates that ventilation has stopped. This can be due to: * An empty O2 cylinder or closed cylinder valve: Check the gas supply in the O2 cylinder and connect to a full O2 cylinder. Open the cylinder valve. * A gas probe not properly connected to the medical gas pipeline system: Check the gas pressure in the medical gas pipeline system to ensure it is greater than 2.7 bar. Fully insert the gas probe into the terminal unit.