20 Supplement to the instructions for use Polaris 600
Troubleshooting
Fault – Cause – Remedy
LEDs at the light control panel of the gimbal mounting
A Status LED
Fault Cause Remedy
Light cannot be switched on.
LED status indicators at the trolley do
not light up.
Power cable not connected. Connect mains power supply.
Battery not charged. Charge battery.
Fuse faulty. Have fuse replaced by experts.
Battery or charger faulty. Contact DrägerService.
Battery is not charging. Power cable not connected. Connect mains power supply.
Fuse faulty. Have fuse replaced by experts.
037
A
LED Possible status Cause
Status LED (A) Permanently orange Light is in battery operation, connect mains power
supply.
If the Power LED remains orange, inform in-house
technician.