22
Care
The instrument does not need any special care.
Dirt and deposits can be removed from the instrument by washing it with
cold water. A sponge can be used for wiping if necessary.
Carefully dab dry the instrument using a cloth.
Maintenance
The instrument should be inspected and maintained annually by suitably
qualified persons (consult: EN 60079-29-2 – Guide for the selection, installation,
use and maintenance of apparatus for the detection and measurement of
combustible gases or oxygen; EN 45544-4 – Electrical apparatus used for the
direct detection and direct concentration measurement of toxic gases and
vapours - Part 4: Guide for selection, installation, use and maintenance and
national regulations).
Recommended calibration interval for measuring channels O2, H
2
S, H
2
and CO:
6 months.
Recommended calibration interval for the measuring channel IR Ex/CO
2
: 12
months.
Calibration interval of other gases: see Instructions for Use for the respective
DrägerSensors.
CAUTION
Abrasive cleaning implements (brushes etc.), cleaning agents and cleaning
solvents can destroy the dust and water filters.
Disposing of the instrument
Dispose of the product in accordance with the applicable rules and regulations.
WEEE
Battery disposal
In accordance with EU Directive 2002/96/EC, this product must not be
disposed of as household waste. This is indicated by with the adjacent
icon.
You can return this product to Dräger free of charge. For information
please contact the national marketing organisations and Dräger.
In accordance with EU Directive 2006/66/EC, batteries and
rechargeable batteries must not be disposed of as household waste
but must be taken to battery collection centres. This is indicated by the
adjacent icon.
Collect batteries and rechargeable batteries as specified by the
applicable regulations and dispose of at battery collection centres.