ENGLISH
The dimensions given in the above table may deviate by 2~3cm depending on
production conditions.
The sleeve is defined as the mounting part in IEC 60601-1.
4-5. Appearance and names and functions of buttons
4-5-1. LEG Pressure Button
Set the pressure to be delivered.
This value is not able to be changed during operation.
Basic pressure is set at 40mmHg This pressure setting can be changed at 10mmHg
Increments. (ex. 40 50 60 20 30 40 mmHg)
The device automatically selects the sleeve when the START/STOP button is selected.
The device has an automatic gradient pressure application which applies gradient
Pressure into each chamber of a sleeve sequentially. (±5mmHg)
4
3
1
2
5
6
Leg
pressure
button
Interval
button
Mode
button
LCD display
Foot
pressure
button
START/STOP
button
25