EasyManua.ls Logo

DS Maref DVT-2600 - Chapter 9: EMC Information; EMC Test Standards and Requirements

DS Maref DVT-2600
144 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ENGLISH
Information on EMC
8
Phenomenon
Basic EMC
standard or test
method
Operating
mode
Port tested Test Voltage
Test level/
requirement
Mains terminal
disturbance
voltage
CISPR11:2015 Operating AC Mains
AC 120 V, 60 Hz
AC 220 V, 60 Hz
AC 230 V, 50 Hz
Group1, Class A
Radiated
disturbance
CISPR11:2015 Operating Enclosure
DC 11.25 V
(Battery)
DC 10.89 V
(Battery)
AC 120 V, 60 Hz
AC 220 V, 60 Hz
AC 230 V, 50 Hz
Group1, Class A
Harmonic Current
Emission
EN 61000-3-2:2014
IEC 61000-3-
2:2014
Operating AC Mains AC 230 V, 50 Hz Class A
Voltage change,
Voltage
fluctuations and
Flicker Emission
EN 61000-3-3:2013
IEC 61000-3-
3:2013
Operating AC Mains AC 230 V, 50 Hz
Pst: 1
Plt: 0.65
Tmax:0.5
dmax: 4%
dc: 3.3%
Electrostatic
Discharge
Immunity
EN 61000-4-2:2009
IEC 61000-4-
2:2008
Operating
Stand by
Enclosure
DC 11.25 V
(Battery)
DC 10.89 V
(Battery)
AC 120 V, 60 Hz
AC 220 V, 60 Hz
AC 230 V, 50 Hz
± 8 kV/Contact
± 2, ± 4, ± 8,
± 15 kV/Air
Radiated RF
Electromagnetic
Field Immunity
EN 61000-4-
3:2006+A2:2010
IEC 61000-4-
3:2006
+A2:2010
Operating
Stand by
Enclosure
DC 11.25 V
(Battery)
DC 10.89 V
(Battery)
AC 120 V, 60 Hz
AC 220 V, 60 Hz
AC 230 V, 50 Hz
3 V/m
80 MHz-2.7 GHz
80% AM at 1 kHz
Immunity to
Proximity Fields
from RF wireless
Communications
Equipment
EN 61000-4-
3:2006+A2:2010
IEC 61000-4-
3:2006
+A2:2010
Operating
Stand by
Enclosure
DC 11.25 V
(Battery)
DC 10.89 V
(Battery)
AC 120 V, 60 Hz
AC 220 V, 60 Hz
AC 230 V, 50 Hz
Table 9 in
IEC 60601-1-2:
2014
Electrical Fast
Transient/Burst
Immunity
EN 61000-4-4:2012
IEC 61000-4-
4:2012
Operating
Stand by
AC Mains
AC 120 V, 60 Hz
AC 220 V, 60 Hz
AC230 V, 50 Hz
± 2 kV, 100 kHz
repetition frequency
Surge Immunity
EN 61000-4-5:2014
IEC 61000-4-
5:2014
Operating
Stand by
AC Mains
AC 120 V, 60 Hz
AC 220 V, 60 Hz
AC 230 V, 50 Hz
Line to Line
± 0.5 kV, ± 1 kV
Line to Ground
± 0.5 kV, ± 1 kV, ± 2
kV
39