4-5-8. Error Description
If there is an error with the device; the error mode will start with an alarm.
The device will revert back to Waiting Mode when the user presses START/STOP.
Error codes and instructions to follow.
1) Code 10 (Sleeve connection error)
During power on the device performs an automatic one cycle calibration to detect the
sleeve type and proper garment fit. If the device fails to detect the sleeve, it shows Error
code 10 with an alarm sound.
*Checking point
Check the connection of the device to the tubings
(tubing).
Check that each sleeve is wrapped properly on the
patient’s leg.
Segment Error information Segment Error information
Sleeve connection
error
This error mark will be
on the screen if there is
no sleeve connection to
the device.
Vent error
Pressure Sensor Error.
Pump error
When the device
pump has an error.
Device error
When the device has
an error.
Low pressure error
Under the 30% pressure
has consecutively
checked three times.
Warning of low battery
Connect the AC power.
High pressure error
Over the 30%
pressure
has consecutively
checked three times.
Error Code
Error 30
(System
error)
Error 10
(Sleeve
connection
error)
Error 20
(Pressure
error)
Error 40
(Power
error)
30