The following information only applies to the V300S�
15 Information about EMC in accordance with EN 60601-1-2
15.1 General information
The information in this leaflet includes excerpts from the relevant European standards for electrical,
medical devices� It must be observed when installing Dürr Dental devices or combining them with
products of other manufacturers� If you are uncertain about anything, please refer to the complete
standard�
15.2 Abbreviations
EMC Electromagnetic compatibility
HF High frequency
U
T
Rated voltage of the device (supply voltage)
V
1
, V
2
Compliance level for the test in acc� with IEC61000-4-6
E
1
Compliance level for the test in acc� with IEC61000-4-3
P Rated power of the transmitter in Watts (W) in accordance with the specifications of the
transmitter manufacturer
d Recommended safety distance in metres (m)
15.3 Guidelines and manufacturer's information
Electromagnetic emissions for all devices and systems
The device is designed for operation in an electromagnetic environment as specified below� The cus-
tomer or operator of the device should ensure that the device is operated in such an environment�
Interference emission
measurements
Compli-
ance
Electromagnetic environment - guidelines
HF emissions in accord-
ance with CISPR11
Group 1 The suction unit uses HF energy exclusively for internal
functions� For this reason, HF emissions are very low
and it is unlikely that any interference will be caused to
neighbouring electronic devices�
HF emissions in accord-
ance with CISPR11
Class B The suction unit is suitable for use in all facilities includ-
ing those in living areas and areas that are directly con-
nected to the public mains electricity supply that also
supplies buildings used for residential purposes�
Harmonics in acc� with
IEC61000-3-2
Class A
Voltage fluctuations/flickers
in acc� with IEC61000-3-3
Compliant
Appendix
EN

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Durr Dental V 1200 S and is the answer not in the manual?

Durr Dental V 1200 S Specifications

General IconGeneral
ModelV 1200 S
ManufacturerDurr Dental
Number of surgeries12
Voltage230 V
Frequency50/60 Hz
Power consumption2.4 kW
Air flow rate per treatment unit300 l/min
Dimensions (H x W x D)690 x 540 x 670 mm
Weight71 kg
Number of Pumps1