INSTALLATION
  1.  LOCATION: Select a location where the temperature of the unit will be between 
      20°F and 120°F. Distance from the receiver is limited only by total loop resistance. 
      See “Electrical Connections”. The tubing feeding pressure to the instrument can 
      be run practically any length required but long lengths will increase response time 
      slightly. Avoid surfaces with excessive vibration.
  2.  POSITION: All standard models are calibrated with the diaphragm vertical and 
      should be used in that position for maximum accuracy. If your application requires 
      mounting in other than a vertical position, be sure to specify this when ordering.
  3.  PRESSURE CONNECTIONS: For convenience, two sets of 1/8˝ female NPT 
      ports are available. Plug the unused set with pipe plugs provided. Attach tubing 
      from positive pressure source to port marked “HI” or from negative (Vacuum) 
      source to port marked “LOW”. In either case, opposite port must be vented to 
      atmosphere. In dusty environments, we recommend use of an A-331 Filter Vent 
      Plug to keep interior of instrument clean. For differential pressures the higher 
      source is connected to the “HI” port and lower to the ”LOW” port.
  4.  MOUNTING: The Series 605 Transmitter may be either panel mounted or 
   surface mounted.
 A.  PANEL MOUNTING: Cut a 4-3/4˝ or 120 mm dia. hole in panel and insert 
            the complete unit from the front. Slip on the mounting ring and install the 
            split snap ring in the groove on the bezel. Seat the mounting ring against the 
            snap ring and thread the four screws through the tapped holes. Tighten 
      screws against rear of panel.
 B.  SURFACE MOUNTING: Drill (3) 3/16˝ dia. holes for mounting screws and 
            cut (1) 7/8˝ x 1-1/8˝ hole for access to terminal strip as shown in hole 
            location drawing. Insert screws from rear of panel and thread into tapped 
            holes on back of transmitter case. If rear pressure connections are to be 
            used, make 1/2˝ dia. holes located as shown in hole location drawing in left 
      column.
 
   C. FOR -SS BEZEL INSTALLATION: Provide a 4-9/16˝ opening in panel. 
            Insert gage and secure with supplied mounting hardware.
  5.   ZEROING: Once gage/transmitter is mounted in its nal position, check to be 
      sure pointer aligns with zero on scale, when no pressure is applied and both low 
      and high pressure ports are vented to atmosphere. To adjust, turn small slotted 
      screw at center-bottom of gage face. Do not move the larger black knobs labeled 
      SPAN and ZERO. These are for use only if a calibration check shows the 4-20 
      mA  output  signal  to  need  adjustment.  See  page  3  under  heading  OUTPUT  
      RANGING.
PNEUMATIC PRESSURE T
17/32
1-7/8
7/8
1-1/8
11/16
21/32
1-3/4
(3) 3/16 DIA. HOLES EQUALLY
SPACED ON A 4-1/8 DIA. BOLD
CIRCLE FOR SURFACE 
MOUNTING
1/2 DIA. HOLE IN PANEL FOR HIG
PRESSURE BACK CONNECTION
WHEN SURFACE MOUNTED
1/2 DIA. HOLE IN PANEL FOR
LOW PRESSURE CONNECTION
WHEN SURGACE MOUNTED
HOLE IN P
ANEL
TERMINAL BLOCK
1-1/8˝ LONG X 
7/8˝ WIDE FOR 
TERMINAL
 STRIP
3/16 DIA. HOLES
FOR MOUNTING SCREWS
TYP. 3 PLACES
1/2 DIA. HOLES FOR 
AND/OR MOUNTING