EasyManuals Logo

E Instruments E8500 Plus User Manual

E Instruments E8500 Plus
49 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
E8500 Instruction & Operations Manual
version 1.017
Page 45 of 49
SENSORS
1. EMISSIONS SENSORS – ELECTROCHEMICAL
SENSOR
RANGE
RESOLUTION
ACCURACY
CARBON
MONOXIDE (CO)
Low Range 0-8000 ppm 1 ppm <300ppm, 10ppm
To 8000ppm, 4%
CARBON
MONOXIDE (CO)
Dilution
Auto-Range
4000-20000
ppm
1 ppm >2000ppm, 10%
NITRIC OXIDE
(NO)
Std. Range 0-5000 ppm 1 ppm <100ppm, 5ppm
To 5000ppm, 4%
NITRIC OXIDE
(Low NO)
Low Range 0-100.0 ppm 0.1 ppm <50.0ppm, 1.5ppm
To 100.0ppm, 4%
NITROGEN
DIOXIDE (NO
2
)
Std. Range 0-1000 ppm 1 ppm <100ppm, 5ppm
To 1000ppm, 4%
NITROGEN
DIOXIDE(Low NO
2
)
Low Range 0-100.0 ppm 0.1 ppm <50.0ppm, 1.5ppm
To 100.0ppm, 4%
SULFUR DIOXIDE
(SO
2
)
Std. Range 0-5000 ppm 1 ppm <100ppm, 5ppm
To 5000ppm, 4%
SULFUR DIOXIDE
(Low SO
2
)
Low Range 0-100.0 ppm 0.1 ppm <50.0ppm, 1.5ppm
To 100.0ppm, 4%
HYDROGEN
SULPHIDE (H
2
S)
Std. Range 0-500 ppm 1 ppm <100ppm, 5ppm
To 500ppm, 4%
2. EMISSIONS SENSORS – PHOTO-IONIZATION
SENSOR
RANGE
RESOLUTION
ACCURACY
VOLATILE
ORGANIC
COMPOUNDS
(VOC)
Std. Range 0-200 ppm 1 ppm 10% rdg. + 1 ppm
3. EMISSIONS SENSORS – NON-DISPERSIVE INFRARED (NDIR) SENSORS
SENSOR
RANGE
RESOLUTION
ACCURACY
HYDROCARBONS
(HC or CxHy)
0 - 0.40 %
0.40 – 1.00 %
1.00 – 3.00 %
0.01 % 3% of Rdg.+ 0.01%
5% of Rdg.
8% of Rdg.
CARBON
MONOXIDE (CO)
High Range
0%-10.00%
10.01%-15%
0.01%
0.02% or 3% Rdg.
5% of Rdg.
CARBON DIOXIDE
(CO
2
)
0.0% - 20.0%
20.0% - 50.0%
0.1% 0.3% or 3% Rdg.
5% of Rdg.