1.1 FEATURES
Separate, economical, easily replaceable Nitrate, Chloride, and optional pH electrodes
Fast and Accurate Nitrate Measurement NO
3
or NO
3
-N
Automatic compensation for Cl interference
Temperature compensated NO
3
-N measurement
Rugged PVC design with removable electrode guard for easy maintenance
Integral Spray Head Cleaner
o Cleans sensor in situ with turbulence caused by pressurized air
Internal Signal Conditioning allows up to 200 meters between Sensor and Analyzer
1.2 HYDRA-DS SENSOR SPECIFICATIONS
Three Electrode System with spray cleaner
Nitrate ISE (NO
3
-N) is the primary measurement
Chloride ISE is used for compensation of the NO
3
signal
Optional pH or NH
4
+
electrode
The Sensor is waterproof with an ingress rating of IP68
1.2.1 MEASUREMENT RANGE
NO
3
-N
Cl
pH
NH
4
+
1.2.2 OPERATING TEMPERATURE
0 °C to 40 °C (32 °F to 104 °F)
1.2.3 MIN/MAX FLOW RATE BY THE SENSOR
Minimum 0.1 m/s
Maximum 3.0 m/s
1.2.4 WETTED MATERIALS
PVC, PEC, PVDF, PTFE, Viton, Glass, 316 SS
1.2.5 ACCURACY
± 3% of reading, dependent on Calibration
1.2.6 RESPONSE TIME
T90 approximately 1 minute
1.2.7 ELECTRODE LIFE
ISEs: 4-6 months, typical
pH electrode: 6-12 months, typical