E C O P H Y S I C S Operating Guide
CLD 780 TR / July 2000 74
(59) “Enter Value!“ window
Enter calibration factors manually.
“Enter Value!“: The user may manually alter the three calibration factors.
Enter a value between 0 and 9999 for “ZEROnet“ and “fNO“ , and be-
tween 0 and 2 for “
λ
NO“ .
Press “YES“ to confirm your entries. These values are now current.
Press “YES“ and “MEAS“ to return to measure mode.
(60) “Exit Calibration“ window.
Abort calibration.
“Exit Calibration Y/N
Press “YES“ to abort calibration and return to measuring mode. The
previous calibration factors remain valid. Window (1a) appears.
Press “NO“ to return to the start of the calibration procedure, window
(51).
Caution: the previous values are now overwritten and have been lost.
Be aware that the instrument makes no plausibility checks on these
values, and senseless combinations may be manually entered and will
be accepted without question by the instrument. The operator bears
full responsibility for the quality of the measured data when he or she
modifies these calibration parameters.