EasyManua.ls Logo

Ecotech A Series

Default Icon
119 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EC 9830
Quick Start Guide.
Step 1 – Installation:
Inspect analyzer for damage before turning on. Service Manual:- 1.1.
Select an appropriate location. Operation Manual:- 2.1.1.
Connect Gas lines. Operation Manual:- 2.1.2.2.
Connect Analog Output Cables. Operation Manual:- 2.1.2.1.
Connect RS232 Cables. Operation Manual:- 4.2.1.
Check the mains power selection switch (115 or 230 VAC). Operation Manual:- 2.2.
Connect AC Mains Power. Operation Manual:- 2.2.
Step 2 – Start-up:
Set Service Switches. Service Manual:- 1.1.2.
Turn On power. Operation Manual:- 2.2.
The Display should read “9830 CO Analyzer”.
Adjust the Display Contrast if required. Operation Manual:- 2.2.1.
Verify that the software is running by observing the Ecotech Globe rotating in the bottom left
hand corner of the display.
Step 3. – Operation:
Verify Instrument warm up and operation mode. Service Manual:- 2.2.
Set the correct time and date. Operation Manual:- 2.3.3.
If using RS232, configure the Interface menu. Operation Manual:- 2.5.12.
Check SYSTEM FAULTS menu. All PASS. Operation Manual:- 2.5.21.
Verify other menu settings. Service Manual:- 4.2.
Step 4. – Calibration:
Perform a quick (single point) calibration. Operation Manual:- 2.4.
Setup and Calibrate the Analog Outputs (if applicable). Operation Manual:- 2.6.3.
If necessary, perform a leak check. Service Manual:- 3.3.10.
If necessary, perform a flow calibration. Service Manual:- 3.5
If necessary, perform a Multipoint calibration. Operation Manual:- 3.2.
Step 5. – Data Validation:
Verify the results from your data acquisition system agree with the readings of the EC9830
CO analyzer.
Verify that the analyzer responds to automatic calibration sequences.
The analyzer is now operating correctly.

Table of Contents

Related product manuals