SE-601 Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 43 -
5.3.4 Power Supply Board Test
To verify the power supply board,
1) Open the main unit following the procedures described in Chapter 6, Electrocardiograph
Disassembly.
2) Switch on the device.
3) Measure the voltage to earth of the test points listed below by using a multimeter.
Item
Test Point
Reference Result
12V
+12V±0.6V
5V
+5V±0.25V
8.3V
8.3V±0.3V
GND
4) Compare the measurement results with the reference results in the list. If any one of the
results exceeds the reference range, the power supply board defection is confirmed.
Replacement of the power supply board is recommended.
12V
5V
GND
8.3V