See illustration for set up from the wall outlet to power supply at the bottom of power adapter (3)
as per the illustration of figure 15-9.
Please take this opportunity to review the EV1000A clinical platform configurations at your site to
ensure they are configured as instructed in the Operator’s Manual, to minimize the risk of fluid
ingress.
For EV1000NI clinical platform installations:
We’ve received reports where liquid is dripping on to the unit which may lead to an electrical
short circuit leading to a possible fire hazard due to liquid ingress. Unlike the EV1000A, the
EV1000NI AC cord inserts horizontally into the unit. Therefore, orientation is not a factor of liquid
ingress. However, in case of excessive liquid exposure, the unit may short circuit leading to a
possible fire hazard due to liquid ingress at the AC cord.
We strongly advise you to be cautious that IV bags and other fluids need to be positioned away
from AC cord in order to avoid liquid ingress.
Please complete the enclosed response form so we may follow up with any further actions as
necessary.
Details of steps 1, 2, 3
are included in the
Operator’s Manual

Questions and Answers: