Applying the finger cu
1. Select the proper size ClearSight finger
cu by using the ClearSight finger cu
sizing aid.
2. Place the middle phalanx of the patient’s
pointer, middle, or ring finger onto
the cu and gently lead the cu cable in
between two finger to the back side of the
hand. The cu must be lined up between
the first and second knuckles (1).
3. Line the middle phalanx of the finger between the two green lines
on the inside of the cu.
4. Wrap the ClearSight finger cu tightly around the finger. Do not rotate
the ClearSight finger cu after application (2).
5. Gently lead the cu cable and air hose between two fingers toward the
back of the hand (3).
6. Connect the ClearSight finger cu to the pressure controller (3).
7. If using a second finger cu, repeat steps 1-4 on adjacent
finger with cu #2.
Applying the heart reference sensor (HRS)
1. Connect the HRS to the pressure
controller (4).
2. Attach the finger end of the HRS to
a single ClearSight finger cu (5).
3. Attach the heart end of the HRS to
the patient at the phlebostatic axis (6).
NOTE: Changing body position during the measurement
may require reapplication of the heart end of the HRS.