Introduction
1.4.4 Analogue output
The Podule produces an analogue output that allows for monitoring four different system parameters:
 Measured pump rotational speed
 Measured motor power
 Measured motor temperature
 Measured controller temperature
The analogue output signal ranges from 0 to 10 V and is directly proportional to the system parameter (refer to 
Section 2.8). 
Connect the analogue output to a suitable meter or indicator to display the appropriate system parameter or connect 
to the customer control equipment (for example, to operate other components in the pumping system at pre-set 
values).
Only one system parameter can be monitored at a time using the analogue output. However, it is easy to configure 
the Podule to monitor a different system parameter (refer to Section 4.1.7).
1.4.5 Automatic vent options
An Edwards TAV vent valve can be connected directly to the DX pump’s Podule. The Podule is capable of controlling 
a number of different venting options. 
The drive electronics can control the rate of venting. Using this feature, the pump can be vented from full rotational 
speed in a controlled manner that will not damage the pump bearings. Once the pump rotational speed has dropped 
to below 50% of maximum speed, it is safe to hard vent (open the vent valve fully.)
There are many venting options available, including:
 Hard vent when rotational speed drops below 50% 
 Controlled vent when above 50% speed and hard vent below 50% speed 
 Hard vent immediately through a suitable restrictor
A full list of the venting options is given in Section 3.8.
In addition, there is a feature that allows a delayed start of the DX pump. With this feature, the vent valve can be 
closed before starting the DX pump. This allows the backing pump to reduce the pressure in the vacuum system 
before starting the DX pump.
If the Podule is not used to control a TAV vent valve, it can be used to run a fan instead. Configure the Podule so that 
the fan is permanently enabled.
1.4.6 Normal speed setting
The Normal Speed is a user-selectable parameter that can be set anywhere from 50% to 100% of full rotational speed. 
When the pump reaches Normal Speed, a signal is available on the Normal pin of the logic interface connector. Use 
this signal to control the application since it shows that pump speed, and therefore vacuum performance, has 
reached a minimum specific level. The default setting is 80% of full rotational speed. Refer to Section 4.1.5 for 
instructions on altering the Normal Speed setting.