6.4 System operang t
emperature conguraon
The system operang temperature conguraon can be adjusted to suit process
demands. Contact your supplier for instrucons on how to access the conguraon
menu (through the PDT) and for process specic recommendaons.
Use the following procedure:
1. Access the Cong menu on the PDT (requires an access code, contact supplier).
2. Scroll to Temp Set-points - press Enter.
3. Scroll to Set DP Temp - press Enter.
4. Scroll to desired temperature - Press Enter.
Note:
The Dry Pump se
t-point is the desired surface temperature of the internal low
vacuum stage.
Refer to Table: Dry pump and booster temperature sengs for default temperature
sengs and ranges.
Table 32
Dry pump and booster temperature sengs
Temperature Seng
Range in which temperature can be
set(°C)
De
fault value (°C)
Dry pump
(Standard variants)
iXM600, iXM1200
60 - 110 80
Dry pump
(Standard variants)
iXM200, iXM3000
60 - 110 110
Dry pump
(T and TX variants)
60 - 130
130
Booster
(Only TX variants)
- 85
10/2020 - ©Edwards Limited
Page 65M56635880_D
M56635880_D - Operaon