EasyManuals Logo

EKO CORE User Manual

EKO CORE
23 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
Guidance and Manufacturer’s Declaration - Electromagnetic
Immunity
The Eko Electronic Stethoscope System is intended for use in the
electromagnetic environment specified below. The user of the Eko
Electronic Stethoscope System should assure that it is used in such an
environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment-
Guidance
Electrostatic
Discharge
(ESD) IEC
61000-4-2
+/- 6 kV
contact
+/- 8 kV
+/- 6 kV
contact
+/- 8 kV air
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%
Electrical Fast
Transient/
Burst
IEC 61000-
4-4
+/- 2 kV
for supply
lines
+/- 1 kV
for input/
output
lines
Not
Applicable
Surge
IEC 61000-
4-5
+/- 1kV
line(s) to
line(s)
+/- 2 kV
line(s) to
earth
Not
Applicable
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-
4-11
< 5% U
T
(>95% dip
in U
T
) for
0.5 cycle
40% U
T
(60% dip
in U
T
) for 5
cycle
70% U
T
(30% dip in
U
T
) for 25
cycle
< 5% U
T
(>95% dip
in U
T
) for
5 sec
Not
Applicable
20

Other manuals for EKO CORE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EKO CORE and is the answer not in the manual?

EKO CORE Specifications

General IconGeneral
TypeDigital Stethoscope
ConnectivityBluetooth
AmplificationUp to 40x
SoftwareEko App
FDA ClearanceYes
CompatibilityiOS, Android
ChargingUSB charging
FeaturesNoise cancellation
Frequency Response20Hz - 2kHz

Related product manuals