8.1 Electrical Safety
Warning: The use of accessories other than those specified, with the exception
of accessories sold by Eko as replacement parts, may result in increased
emissions or decreased immunity of the Eko Electronic Stethoscope System.
Warning: The Eko Electronic Stethoscope System should not be used adjacent
to or stacked with other equipment. If adjacent or stacked use is necessary,
the Eko Electronic Stethoscope System should be observed to verify normal
operation in the configuration in which it will be used.
Guidance and Manufacturer’s Declaration - Electromagnetic
Emission
The Eko Electronic Stethoscope System is intended for use in the
electromagnetic environment specified below. The user of the Eko
Electronic Stethoscope System should assure that it is used in such an
environment.
Applicable
Emissions Test
Compliance Electromagnetic
Environment- Guidance
RF emissions
CISPR 11
Group 1 The Eko Electronic Stethoscope
System uses RF energy only for
its internal function. Therefore,
its RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B The Eko Electronic Stethoscope
System is suitable for use in
all establishments, including
domestic establishments and
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
Harmonic Emissions
IEC 6100-3-2
Not Applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not Applicable